Schott SI Analytics
On demand
Schott SI Analytics ScienceLine Metal Combination Electrode AgCl 6280 with Coated Sensor, 2 models. The electrode, not the meter, is what decides whether a read
285102351
285102413
Schott SI Analytics ScienceLine Metal Combination Electrode AgCl 6280 with Coated Sensor. The electrode, not the meter, is what decides whether a reading is right.
The models differ by 6280 and 62.
An electrode ages from the day it is unpacked, whether or not it is used. The glass membrane hydrates and eventually degrades, the reference junction clogs with whatever it has been dipped in, and the internal electrolyte is consumed. A meter that reads badly is far more often carrying a tired electrode than a fault.
Budget for replacement on a schedule and keep the calibration record: a slope drifting away from 100% and an offset drifting away from zero are the two numbers that say when.
The junction is what matters most. A general-purpose ceramic junction suits clean aqueous samples and clogs quickly in anything else. A sleeve or flowing junction survives slurries, proteins, tris buffers and plating baths. A flat surface tip reads a solid — paper, skin, cheese, soil — where a bulb cannot make contact at all.
Store the tip wet in storage solution — never in distilled water, which strips the reference electrolyte and shortens the electrode's life.
Supplied in 2 models.